The aim of all extraction procedures is to
separate cellular or fluid lipids from the other constituents, proteins, polysaccharides,
small molecules (amino acids, sugars...) but also to preserve these lipids for further
analyses. The preservation of proteins involves special precautions found in delipidation
procedures.
There is a great diversity of methodologies because biological tissues are not similar
when considering their structure, texture, sensitivities and lipid contents. Removing the
non-lipids without losing some lipids is a complex challenge, extracting some specific
lipids is not always reliable for other kinds of lipids.
The high sensitivity of the analytical methods needed for low amounts of extracted lipids
requires the use of very pure solvents and clean glassware. The chemical nature of the
extracted lipids must also be taken into consideration.
Furthermore, all lipids must be protected against degradation through oxidation by
solvent, oxygen, enzymes in combination with temperature and light.
If you want to learn more on lipid extraction :
If we describe here the extraction on a laboratory scale of
lipids from various biological materials to run analytical procedures, different
processes may lead also to the extraction of oil from many seeds, nuts, and
kernels for use as nutritional supplement, as well as raw material for
industrial applications.
The type of instrument that is appropriate depends on the size of the operation.
Oil processing operations range from "cottage industries" processing
some kg per day, to factories processing several thousand tons of seed per day.
In all cases, the sequence of operations includes cleaning, dehulling, grinding,
and pressing. Furthermore, oils generally need refining to remove cloudiness,
excess color and unpleasant flavors. As many crude oils contain variable amounts
of mucilaginous compounds (gums) combined with phospholipids, a degumming
process has to be run (Dijkstra AJ, OCL 1998, 5, 367). In several instances, and especially when
seeds have low oil content (like soybeans) more complex methods including
solvent extraction are processed. If hexane is the most commonly used solvent,
isohexane appears to be the most likely candidate to replace n-hexane as the
preferred oilseed extraction solvent, mainly in USA to avoid federal
environmental regulations applying to the use of n-hexane. Several studies have
shown that the extraction performances of the two isomers are quite similar (Inform
2002, 13, 282).
Those who are interested in oilseed processing will find a comprehensive
description of basic processes with sources for additional information and a
list of suitable raw material in the ATTRA
(National Sustainable Agriculture Information Service) publication :
"Small-scale oilseed processing. Value-added and processing guide"
Aqueous enzymatic oil extraction is an emerging technology since
it offers many advantages such as elimination of solvent consumption, degumming
operations and removal of some toxins. A review of aqueous and enzyme-based
processes may be consulted (Rosenthal A et al., Enzyme Microb Technol 1996,
19, 402). Applications of that methodology to the oil extraction of Irvangia
seed kernels (Women HM et al., Eur J Lipid Sci technol
2008, 110, 232) and canola seeds (Latif S et al., Eur J Lipid Sci Technol
2008, 110, 887) have been reported.
Comparative experiments on tobacco seeds have demonstrated that the best oil
recovery was obtained by cold pressing (Stanisavljevic IT et al., Eur J Lipid
Sci Technol 2009, 111, 513). Shortly, seeds were pressed by a hydraulic
press (Komet,
Germany) through three nozzles (diameter of 15, 10 and 6 mm). The press does not
involve mixing and tearing of the seeds. The seed cake was ground by an
electrical mill and the residual oil was extracted by using n-hexane at a
seeds-to-solvent ratio of 1 : 3 g/mL and 25°C for 60 min. This recovery
technique is said to be more acceptable than the other methods, not only for
economic but also for environmental, health and safety reasons.
GENERAL AND PRACTICAL ASPECTS
As carelessness at the preliminary stage of extraction may result in the loss of components of interest or in the production of artifacts, the lipidologist must be aware of various precautions that are summarized below.
Solvents
Glassware
Removal of contaminants
Practical aspects
Storage
![]()
Neutral lipids or generally
storage lipids are extracted with relatively non-polar solvents such as diethyl ether or
chloroform but membrane-associated lipids are more polar and require polar solvents such
as ethanol or methanol to disrupt hydrogen bondings or electrostatic forces.
To avoid peroxidation of the extracted lipids, during the procedure or later, all solvents
should be used peroxide-free. The most convenient is to use "pesticide grade" or
HPLC grade" and to buy small quantities at once. Do not stock bottles for a year !!
Manufacturers often indicate an expiration date on bottles. To minimize possible
deleterious effects due to peroxide formation, pay attention to the expiration dates and
keep stocks for no more than 3 months. Furthermore, to avoid contamination of
extracted lipids by non-volatile material contained in the large volume of solvent used in
the first step of extraction, it is important to use organic solvent containing less than
5 mg dry matter per liter, 1 mg/l is currently proposed.
Diethyl ether, dioxan, isopropyl ether
and tetrahydrofuran form peroxides on storage and must be kept in dark
bottles or in dark areas. These solvents should be routinely tested to prove that
peroxide levels are kept low before use. It must be emphasized that peroxides, even at low
levels, may present an explosion hazard after concentration of solvents. Because
of its low flammability and its natural origin, ethyl lactate was proposed as a
substitute for ethyl acetate in exctracting carotenoids from foods (Ishida BK
et al., J Agric Food Chem 2009, 57, 1051).
Alcohols are good solvents for most lipids, methanol and
ethanol being the most popular. Ethanol can be contaminated on storage by
aldehydes.
Chloroform is a popular solvent, particularly for lipids of intermediate
polarity and when mixed with methanol it becomes a general extraction solvent. It is not
very stable, forming phosgene and HCl in air, it should be stored in brown bottles. It is
sold after addition of preservatives, ethanol, amylene or cyclohexene. Dichloromethane
(or methylene dichloride) is a similar extractant but less oxidizable. Chloroform and
dichloromethane are currently stabilized by addition of ethanol (up to 1%) or
methyl-2-butene-2 (about 20 mg/l).
Among hydrocarbons, hexane is the most popular but is a good solvent only
for lipids of low polarity. Its main use is to extract neutral lipids from mixtures of
water with alcohols. A mixture of isomers, called "hexanes", can be used for the
same purpose. Hexane can be replaced by petroleum ether which is a mixtures of various
hydrocarbons with 5 to 8 carbon atoms. Cyclohexane is sometimes used to
store lipid extracts in the cold without danger of evaporation since it freezes at about
6°C. Benzene is no more used since it is now considered as a potent
carcinogenic substance, it may be replaced by toluene which, nevertheless, is more
difficult to evaporate.
A simple test for peroxides in water-immiscible solvents:
shake 2 ml of solvent with 1 ml of a freshly prepared KI solution (10% in water) and a
drop of starch indicator solution, a blue color is rapidly developing if peroxides are
present (a faint yellow/brown color after addition of KI indicates low levels of
peroxides).
The leitmotiv is to avoid any
source of contamination by mineral oils, greases, plasticisers from plastics and
detergents. Therefore, all operations should be carried out in glass equipment with, if
required, ground-glass joints (do not grease !!). Separatory funnels or columns are best
equipped with Teflon (PTFE) stopcocks.
All vials or tubes should be closed if necessary (for incubation, agitation, centrifugation) with screw caps including a Teflon-covered liner (Teflon inside the tube
!!). Never use cork, rubber, polyethylene or Para film.
Filtration of HPLC solvents should be made with an all-glass vacuum filtration apparatus
equipped with filter membranes made of nylon or Teflon.
Except Teflon, all plastics must be avoided because they leach contaminants into the
solution which can be detected as unknown spots on thin-layer plates or extra-peaks in GLC
chromatograms. Polypropylene tubes can be used if GLC is not necessary for the analysis.
It should be emphasized that all the glassware must be cleaned using classical basic
detergents in washing machine or sonicator bath but the washed vials or tubes should be
first tested for the absence of contaminants with the current techniques.
In the presence of some water, small molecules
are dissolved in organic solvents. Among these contaminants there are pigments,
lipophilic amino
acids, polypeptides and some hydrophobic proteins.
The removal of contaminants can be done by various ways:
- A simple way to eliminate "proteolipids" is to evaporate the lipid extract
under vacuum, add a small amount of methanol which is an excellent denaturation agent and
evaporate again. Redissolve the dried lipid extract with a small volume of
chloroform/methanol (2/1) and transfer into another clean tube.
- Use the washing procedure of Folch's method and repeat the process if necessary with the
same volume of the reconstituted upper phase. Finally, a chloroform phase can be washed
efficiently with water by vortexing and centrifugation.
- Use a liquid/liquid partitioning column. The Sephadex packing (G 25 superfine) is
swollen (20 g/250 ml of solvent) in a mixture of chloroform/methanol/water (60/30/4.5) and
washed three times (30 min each time). A small amount of suspension is poured in a small
glass column (1 cm diameter, 2.5 cm height for filtration of up to 10 mg lipids). The
lipid sample is dissolved in the same solvent mixture and 2.5 ml are applied to the column
followed by 5 ml of the same mixture, 2.5 ml of chloroform / methanol (2/1) and finally by
2.5 ml of chloroform / methanol / water (48/35/10). The total purified lipid extract (12.5
ml) is evaporated under a stream of nitrogen and redissolved in the required amount of
solvent. This procedure is commonly used for gangliosides studies where colored
contaminants frequently hide lipid spots on HPTLC (Dreyfus et al Anal Biochem 1997, 249,
67).
Alternatively, all lipids except gangliosides are separated from various
impurities (bile salts, amino acids sugars) on a Sephadex G-25 column previously
equilibrated with chloroform/methanol (19/1, v/v) saturated with water. A glass
column with a Teflon stopcock and a fiberglass plug is used. The Sephadex beads
are equilibrated in 1/1 methanol/water and added to a height of 10 to 20 cm.
Five cm of washed sea sand is then added on the top. Lipids dissolved in
chloroform/methanol (19/1, v/v) are
applied on the column and eluted with 50 ml of the same solvent. The column is
regenerated with 1/1 methanol/water and then with 19/1 chloroform/methanol
saturated with water and thus can be used indefinitely.
Chlorophyll removal : High levels of chlorophyll-type compounds
are frequently found in lipid extracts from plants. The presence of these
pigments may interfere with further lipid fractionation. Thus, a method has been
developed to chemically remove chlorophyll without modification of other
extracted lipids (Bahmaei M et al., JAOCS 2005, 82, 679).
Crude lipid extracts are mixed with a mechanical device at 200 rpm with a 0.4wt%
mixture of phosphoric and sulfuric acids (2/0.75, v/v) for 5 min at 50°C. The
precipitate is removed from the oily extract by centrifugation (4000 x g, 5 min)
and the oil extract is washed according to the Folch procedure to remove acid
traces.
Glassware: Since all
solvents
may extract also some contaminants from containers or apparatus, only Teflon lined
stoppers and clean glassware must be used. During extraction take care of any contact
between solvent and washers or greased bearings in your mechanical device.
Antioxidant: When fatty acid analyses are planned, it is advisable to add
an antioxidant such as butylated hydroxytoluene (BHT) (about 100 mg per liter).
Lyophilized tissues: They are difficult to extract, thus it is
recommended to rehydrate them with distilled water before extraction (about 5 ml per g dry
material).
With some tissues: it is recommended to add first the required methanol
amount during the homogenization to prevent clogging, this is currently the case with
liver, muscle, blood. chloroform is then added before the last mixing and the true
extraction step.
Evaporation of solvents: When a large amount of solvent
must be evaporated, it should done in a rotary film evaporator, the flask containing the
extract being maintained at no more than 50°C with a water bath. The reduced pressure
must be generated by a pump running without oil (motorized pump with metallic or Teflon
membranes) or conveniently with a water aspirator achieving a vacuum of 10-20 mm Hg. Last
traces of water may be removed by adding 1 or 2 ml ethanol and evaporating again.
The first description of that
concentration device was made by Craig LC (Craig LC et al., Anal Chem 1950,
22, 1462), the famous inventor of countercurrent distribution. The diagram
given in the Craig's paper (see below) shows clearly the principle of the device
found on the market as yet.

If small
samples are extracted, the small solvent volumes (1-5 ml) are easily evaporated under
nitrogen blowing and warming the tubes with the help of a water bath or a heating fan.
With the later device, an excessive heating must be avoided as some free fatty acids or
their esters (C14 and C16) escape from the tubes.
A very valuable and efficient evaporation equipment may be found from Pierce, the Reacti-ThermTM System where reaction vials are heated in a thermostated aluminium
block and the Reacti-VapTM Evaporator module enabling a gentle nitrogen blow.
Lipids should not be left for
any time in the dry state but should be taken up rapidly in an inert non-alcoholic solvent
such as chloroform.
If the weight of extracted lipids are needed, the desiccation time must be prolonged till
a constant weight is reached. A convenient procedure is to evaporate a part of the total
extract in a small flask and to keep it overnight under vacuum before weighing.
Lipid extracts should not be stored in the dry
state. Oxidative degradation is slower in solutions even in the absence of added
antioxidants; This protection depends on the chemical nature of the solvent and the
physical conditions of storage.
Air and light should be avoided. The best storage conditions are a super freezer (-70°C)
to store for a long time (up to one year) lipid extracts in chloroform-methanol mixtures
filling well stoppered glass vials (with Teflon liners), the cap being secured with a wide
length of self-sticking tape. Pencil written labels should be protected by an outer layer
of polyester tape. For long period of storage, it can be useful to flush vials or tubes
with nitrogen before closing to prevent fatty acid oxidation. For the same purpose, a low
amount of antioxidant such as BHT (butylated hydroxytoluene or
2,6-di-tert-butyl-4-methoxyphenol) or ethyl gallate (i.e. about 50 µg BHT/ml) may be added
in the solvent if the natural antioxidant amount is estimated too low (purified extracts).
The antioxidant is used as a concentrated solution in ethanol (10 mg/ml).
After long periods of storage, the cleavage of ester lipids and plasmalogens must be
expected with a concomitant rise in free fatty acids, diacylglycerols methyl or ethyl
esters. These problems can be minimized by using neutral extracts and 2-propanol instead
of methanol or ethanol in the solvent mixture. It must be remembered that best results are
obtained with freshly prepared materials.
To prevent labile lipids from oxygen attack it is possible to use a commercially
available oxygen
absorber (Ageless, Mitsubishi Gas Chem Co) placed inside the sample package.
After a short "pretreatment" at 2°C, the labile lipids appear
preserved even at room temperature for a long time. This method allows
transportation of biomaterial samples to a distant laboratory without utilizing
any freezing system (Hirao S et al., J Oleo Sci 2003, 52, 583).
It has been demonstrated that after one year storage of plasma samples at -80°C
various lipid parameters (cholesterol, triglycerides) were altered (Devanapalli
B et al., Clin Chim Acta 2002, 322, 179). Thus, even at low temperature,
intact biological samples must be extracted as soon as possible. In contrast, it
was shown that after storage at -80°C up to 4 years, the fatty acid
compositions of plasma triglycerides and erythrocyte phospholipids were
practically unchanged (Hodson L et al., Clin Chim Acta 2002, 321, 63).