LIGHT-SCATTERING
AND
CHARGED AEROSOL DETECTORS
The evaporative light detection system has
revolutionized the analysis of lipids by HPLC since its introduction around 1980. This
type of detector works by measuring the light scattered from the solid solute particles
remaining after nebulization and evaporation of the mobile phase. For native lipids (not
derivatized), the light-scattering detector (ELSD) is far more useful for on-line lipid
quantification than the commonly used UV detector.
More recently, an alternative instrument derived from the light scattering
detector was proposed (Corona®
CAD®
instrument from ESA).
This new detection device, charged aerosol detector, is based on an unique technology, in which the HPLC
column eluent is first nebulized with nitrogen and the droplets are dried to
remove mobile phase, producing analyte particles. Then, a secondary stream of
nitrogen becomes positively charged as it passes a high-voltage, platinum corona
wire. This charge transfers to the opposing stream of analyte particles. The
charge is transferred to a collector where it is measured by a highly sensitive
electrometer, generating a signal in direct proportion to the quantity of
analyte present. This device shows consistent inter-analyte response independent
of chemical structure. This means that the Corona detector can be used routinely
for quantitation. This technology is said to be superior to light scattering for
quantitative measurements.
An updated bibliography of papers relevant to various applications of
light-scattering detection to the analysis of lipids is given below :
***********************************
PRINCIPLE OF THE LIGHT-SCATTERING DETECTOR
This type of detector can be used for all
solutes having a lower volatility than the mobile phase. Semi-volatile solutes can be
detected due to the relatively low temperatures at which new generations of instruments
can operate. For the majority of non-volatile solutes, the detection limits reach
frequently the nanogram range per injection.
In the ELSD, the mobile phase enters the detector, is evaporated in a heated device and
the remaining solute is finally detected by the way it scatters light. The intensity of
the light scattered from solid suspended particles depends on their particle size.
Therefore, the response is dependent on the solute particle size produced. This, in turn,
depends on the size of droplets generated by the nebulizer and the concentration of solute
in the droplets. The droplet size produced in the instrument nebulizer depends on the
physical properties of the liquid and the relative velocity and flow-rates of the gas and
liquid stream. The importance of all these parameters emphasizes the need for careful
design and rigorous optimization of the instrument parts.
Three
important steps can be defined during the working of the instrument, steps which are
located in three different parts of the detector: the nebulizer, the evaporation chamber
and the detection system (Figure).
1- Nebulization of the
effluent
This first step transforms the whole liquid
phase flowing from the HPLC column into fine droplets. The larger the droplet size, the
higher the temperature needed to evaporate the liquid phase. The bigger the residual
solute particles, the more intense the scattered light will be.
In general, the nebulization part incorporates a Venturi-type flow of gas around the
eluent inflow. Each brand has its own design, but similar to that found in atomic
absorption spectrometer where fine dispersion are needed. In some instruments, the
nebulization chamber has a special shape and is used to eliminate the biggest droplets of
solvent, thus removing the water part of the sample droplets, in others all the column
effluent passes into the next part (evaporation zone).
2- Evaporation of the effluent
This second step begins when the droplets are
carried by the gas flow into the heated area located before the detection chamber. Each
brand has its own design for this zone and the efficiency of the required evaporation
depends on the shape of the tube and the needed temperature. The solvent is completely
removed to produce particles of solutes without solvation or even droplets of pure
solutes. Practically, a temperature in the range 40-60°C is sufficient to evaporate
solvents used in HPLC of lipids where high percentages of water or polar solvents are
frequently used.
3- Detection
The sample particles pass through a flow cell
where they are hit with an incident light beam, the amount of light scattered being
measured using a photomultiplier and an electronic device. In some instrument a secondary
gas inlet is used to concentrate the particles in the center of the detection chamber.
![]()
CHARACTERISTIC PROPERTIES OF THE DETECTOR
PRACTICAL CONSIDERATIONS
Solvent quality
The solvent quality is of prime importance to
get the lowest background signal. Don't forget that the detector sensitivity is related to
the ratio signal/noise. Don't choose a solvent for ELSD as for UV or fluorescence
detection. The residue after evaporation is the most important criteria and this quality
must be always lower than 1 mg/l. We daily observed that HPLC-grade solvents have this
specification. Solvents used or solvent mixtures must be filtered through compatible
sub-micron filters (0.4 or 0.2 µm). If possible samples must also be filtered with special
filters before injection. When possible, the use of solvents "for residue
analysis" is recommended, they are known to contain less than 5 mg impurities per
liter.
To differentiate the noise from the solvent and that of the instrument, switch off the
pump: you will observe a signal shift for the contribution of the solvent.
Solvent modifiers
A common way to modify the pH value of the
final solvent mixture is to add solvent modifiers. Remember that additives containing
metal cations cannot be used without compromising the ELSD sensitivity. It is possible to
use volatile acids, bases and salts but trials are needed to choose the most efficient
modifier for the separation. Use the lowest modifier concentration compatible with
reliable results. To acidify, use of formic, acetic, trifluoroacetic and nitric acids are
possible. To basify, use ammonia, triethylamine and pyridine. To increase ionic strength,
it is possible to add ammonium bicarbonate or acetate, if necessary in combination with
the above-mentioned acids or bases. Practically, we use no more than 0.1 ml acidic or
basic modifier per liter of solvent.
The response of the ELSD to various lipid species was shown to
be enhanced by the addition of 0.1% (v/v) triethylamine and an equimolar amount
of formic acid in the mobile phase (Deschamps FS et al., Chromatographia
2001, 54, 607). This response modifier enhanced the detection response of
lipids by 2 up to 50 times without altering retention except with zwitterionic
phospholipids which had a better peak shape. It was further demonstrated that
the response enhancement was more marked at low flow rate and was dependent on
solutes and solvents (Deschamps FS et al., Analyst 2002, 127, 35). The
analysis of all parameters led to the conclusion that the chosen modifiers
mainly act as amplifier by the inclusion of triethylamine-formic acid clusters
inside the droplets. A similar effect of these modifiers was observed when they
are added at the outlet of columns eluted with supercritical CO2
(Lesellier
E et al., J Chromatogr A 2003, 1016, 111).
Gas cleanliness
Nitrogen is sometimes preferred for safety
reasons but clear compressed air is most frequently used. In all cases, the instrument
exhaust must be connected to a fume hood. If the air compressor is oil-lubricated, a
charcoal column followed by a sub-micron particle filter must be inserted in the air line
before entering the ELSD. Furthermore, we have inserted after the charcoal column a
dessicator-containing column to decrease the hygrometry of the delivered air.
Internal standards
To ensure accuracy and precision in the
analysis of lipid components with the ELSD, as with other detectors, a calibration against
well-defined lipid standards is necessary.
It remains possible to include in the sample a defined and well separated compound as an
internal standard. As an example, cholesterol was used for phospholipid quantification.
This approach allows the correction of detector drift in between day measurements but
after calibration with a known standard mixture corresponding to the separated compounds.
![]()
Several instruments can be
found on the market for light scattering or charged aerosol detection. All are constructed on a common basis but each one
has a different
design and specific features focusing on various analytical aspects.
- Among the available instruments based on light scattering detection, the analyst may look at several companies:
In the CHROMACHEM (EUROSEP), the heated area is a coiled stainless steel tube which
allows evaporation of pure water at up to 4 ml/min at 55°C. This important device was
designed to allow the use of low temperatures in normal run conditions.
In the PL-ELS 1000 (Polymer Laboratories),
new features are found including a high efficiency nebulizer design and low volume/low
dispersion evaporation system (straight evaporation chamber). The instrument is said
to have baseline stability and very good reproducibility and sensitivity.
Other efficient device may be found in various suppliers : Shimadzu,
Alltech,
Sedere, Waters,
...
- Charged aerosol detection may be effected with the CoronaTM
instrument from ESA.
An overview of that technology and its advantages
may be found on ESA web site or Click
here to download an application note describing the Corona CAD's
sensitivity, dynamic range, response from compounds of different structures
including lipids, and
the instrument's reproducibility characteristics.